The list below contains Medical Device’s Field Corrective Action for the month of April 2026. This list provides field corrective action reporting information to MDA by establishment in accordance to requirement stipulated in Section 41, Act 737 and Regulation 6 of Medical Device (Duties and Obligations of Establishment) Regulations 2019.

Field Corrective Action (FCA) is part of a post-marketing risk assessment measure to ensure the continued safe use of medical devices in the Malaysia market. It is a corrective or preventive action in relation to a medical device which may include –

  • The return of the medical device to the establishment;
  • Modification of the medical device;
  • Exchange of the medical device;
  • Destruction of medical device; or
  • Specific advice on the use of the medical device. 

If you have experienced any safety and performance issue with one of the medical devices listed for FCA action, please seek advice from your medical device authorised representative as soon as possible.

      MEDICAL DEVICE FIELD CORRECTIVE ACTION LISTING APRIL 2026

No.

Date Received

Title of FCA

Affected Medical Device

MDA Reference Number

MDA Registration Number

Local Establishment Contact Detail

1

07/04/2026

VITROS™ Chemistry Products Mg Slides May Not Meet On-Analyzer Stability Specification

VITROS CHEMISTRY PRODUCTS ELECTROLYTES REAGENTS

 MDA/FCA/P1733-55180019-2026

GD9627525-205666

T T MEDICAL MANAGEMENT SDN BHD

[email protected]

2

09/04/2026

2024-PD-MR-016 Philips MR system breast coils Potential for harm while using a breast coil

MR 5300

 MDA/FCA/P1735-97001639-2026

GC4777625-217427

PHILIPS MALAYSIA SENDIRIAN BERHAD

[email protected]

3

14/04/2026

Field Safety Notification for Flexitron System - Intrauterine tube (IU tube)

FLEXITRON SYSTEM

MDA/FCA/P1736-94420885-2026

IVDC51660257518

NYPRAX BUSINESS SOLUTIONS

[email protected]

4

09/04/2026

FA1536 SynchroMed Flex Infusion Mode - A810 Clinician Programmer Application

SYNCHROMED II PROGRAMMABLE PUMP SYSTEM AND ACCESSORIES

 MDA/FCA/P1737-84052626-2026

GB7982922-110702

MEDTRONIC MALAYSIA SDN BHD

[email protected]

5

13/04/2026

Field Safety Action on weak or absent staining associated with specific lots of Artisan Periodic Acid-Schiff Stain Kit (AR165), Artisan Alcian Blue/PAS Stain Kit (AR169), Artisan PAS Green Stain Kit (AR172), and Artisan Alcian Blue/PAS/Hematoxylin Stain Kit (AR178)

DAKO_HISTOLOGY/CYTOLOGY_HISTOLOGY/CYTOLOGY REAGENTS_ARTISAN SPECIAL STAIN

 MDA/FCA/P1745-47535127-2026

GC2975220-43026

BITA LIFESCIENCE SDN. BHD.

[email protected]

6

13/04/2026

Voluntary Recall for InZone Detachment System due to Premature Battery Drainage

INZONE DETACHMENT SYSTEM

MDA/FCA/P1746-54168593-2026

GC84604193017

STRYKER CORPORATION (MALAYSIA) SDN. BHD.

[email protected]

7

13/04/2026

MR015/25/P (CAN) - Distribution of a Customer Advisory Notice to inform MR customers that Siemens Healthineers has incorporated a mandatory visual inspection and improved helium leak check of the venting system into the routine annual maintenance procedure, with clear intervention criteria for when to de ice the magnet via established service procedures

MAGNETOM PRISMA

 MDA/FCA/P1747-97632639-2026

GC5206022-82422

SIEMENS HEALTHCARE SDN. BHD.

[email protected]

8

17/04/2026

ETC Beadchip Kit Exposure Test Carrier Low Signal Results in BeadChipTM Assays

PRECISETYPE HEA MOLECULAR BEADCHIP TEST

 MDA/FCA/P1749-41278792-2026

GC13082609918

BMS DIAGNOSTICS (M) SDN BHD

[email protected]

9

17/04/2026

Blood Pressure Monitor YE660E with the situation where the information of the Outdated Labelling, QR Code Discrepancy and Labelling / IFU Compliance

YUWELL ELECTRONIC BLOOD PRESSURE MONITOR

 MDA/FCA/P1750-10997906-2026

GD17535444617

RBD HEALTHCARE SDN BHD

[email protected]

10

20/04/2026

da Vinci X, Xi Systems Cannula Mount Screws Breakage

DA VINCI XI SURGICAL SYSTEM, ENDOSCOPIC INSTRUMENT CONTROL SYSTEM, MODEL IS4000

 MDA/FCA/P1751-61224737-2026

GB18651834218

DTG MEDICAL SDN BHD

[email protected]

11

20/04/2026

Xpert Carba-R incorrect shipment to Malaysia

XPERT CARBA-R ASSAY

 MDA/FCA/P1753-72098743-2026

GC82521235717

RADIOMETER MALAYSIA SDN. BHD.

[email protected]

12

22/04/2026

ABL800 - Cybersecurity vulnerabilities

ABL8XX SERIES BLOOD GAS ANALYZERS

 MDA/FCA/P1757-22640254-2026

GC59959884718

RADIOMETER MALAYSIA SDN. BHD.

[email protected]

13

22/04/2026

ABL90 FLEX & ABL90 FLEX PLUS - Cybersecurity vulnerabilities

ABL90 FLEX

 MDA/FCA/P1758-88425376-2026

GD958991229918

RADIOMETER MALAYSIA SDN. BHD.

[email protected]

14

27/04/2026

Potential failure of the mechanical ventilation – Extension of TSB13 (MDA/FCA/P1007-13171759-2024)

ATLAN

 MDA/FCA/P1760-60719092-2026

GC4930621-68598

DRAEGER MALAYSIA SDN BHD

[email protected]

15

28/04/2026

Olympus to Provide Update on Instruction for Use (IFU).

SINGLE LIGATION DEVICE_HX-400U-30

 MDA/FCA/P1763-30087646-2026

GD69454346817

OLYMPUS (MALAYSIA) SDN. BHD.

[email protected]

16

29/04/2026

2025-IGT-BST-020 - Philips Allura and Azurion systems Deaeration hose of tube cooling unit may degrade over time and develop oil leakage, potentially leading to loss of imaging functionality or exposure of users and/or patients to fumes

AZURION 3

 MDA/FCA/P1764-43880030-2026

GC92659835718

PHILIPS MALAYSIA SENDIRIAN BERHAD

[email protected]

17

29/04/2026

2025-IGT-BST-020 - Philips Allura and Azurion systems Deaeration hose of tube cooling unit may degrade over time and develop oil leakage, potentially leading to loss of imaging functionality or exposure of users and/or patients to fumes

AZURION 7

 MDA/FCA/P1766-75420104-2026

GC98026871418

PHILIPS MALAYSIA SENDIRIAN BERHAD

[email protected]

18

29/04/2026

2025-IGT-BST-020 - Philips Allura and Azurion systems Deaeration hose of tube cooling unit may degrade over time and develop oil leakage, potentially leading to loss of imaging functionality or exposure of users and/or patients to fumes

AZURION BIPLANE

 MDA/FCA/P1767-20693483-2026

GC91655727318

PHILIPS MALAYSIA SENDIRIAN BERHAD

[email protected]

19

29/04/2026

2024-IGT-PUN-009 Zenition 10 Arcing in X-ray tube potentially leading to temporary loss of imaging functionality

ZENITION 10

 MDA/FCA/P1768-95809475-2026

GB4127824-160695

PHILIPS MALAYSIA SENDIRIAN BERHAD

[email protected]

20

30/04/2026

2024-IGT-BST-023 Philips Azurion Systems Mechanical wear of the Float Tabletop control potentially leading to limited availability or loss of longitudinal and transverse table movements and/or lacerations

AZURION 3

 MDA/FCA/P1769-64198925-2026

IVDB7126524-184610

PHILIPS MALAYSIA SENDIRIAN BERHAD

[email protected]

21

30/04/2026

2024-IGT-BST-023 Philips Azurion Systems Mechanical wear of the Float Tabletop control potentially leading to limited availability or loss of longitudinal and transverse table movements and/or lacerations

AZURION 7

 MDA/FCA/P1770-99059540-2026

IVDA9233125-201119

PHILIPS MALAYSIA SENDIRIAN BERHAD

[email protected]

22

30/04/2026

2024-IGT-BST-023 Philips Azurion Systems Mechanical wear of the Float Tabletop control potentially leading to limited availability or loss of longitudinal and transverse table movements and/or lacerations

AZURION BIPLANE

 MDA/FCA/P1771-34505839-2026

IVDA9233125-201119

PHILIPS MALAYSIA SENDIRIAN BERHAD

[email protected]

* The information contained in the Medical Device Authority Field Corrective Action database is released under Regulation 7(8) and Regulation 8(5) of Malaysia’s Medical Device Regulations 2019.